EasyManuals Logo

Flintec DAD143 Series User Manual

Flintec DAD143 Series
95 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #74 background imageLoading...
Page #74 background image
DAD 143.x User Manual Profinet Rev.1.0.0 October 2022 Page 82 of 95
17 Appendix B Parameter list for the DAD 143.x
Index
Sub-
Index
Acyclic
Index
Name
Type
Attri-
but
Default
Value
Meaning
0x2000
0x00
0x01
0x02
0x03
0x04
0x05
0x06
0x0020
0x0021
0x0022
0x0023
0x0024
0x0025
0x0026
Number of entries.
Gross weight.
Net weight.
Tare.
Dosed weight.
Dosed tare.
Average weight.
UI8
REAL32
REAL32
REAL32
REAL32
REAL32
REAL32
ro
ro
ro
ro
ro
ro
ro
0x06
-
-
-
-
-
-
Number of entries.
Get gross weight GG command.
Get net weight GN command.
Get tare weight GT command.
Get dosed weight GD command
(FT1 & FT3 only)
.
Get dosed tare weight DT command
(FT1 & FT3 only).
Get average weight GA command.
0x2001
0x00
0x01
0x02
0x03
0x04
0x05
0x06
0x0040
0x0041
0x0042
0x0043
0x0044
0x0045
0x0046
Number of entries.
Gross weight.
Net weight.
Tare.
Dosed weight.
Dosed tare.
Average weight.
I32
I32
I32
I32
I32
I32
I32
ro
ro
ro
ro
ro
ro
ro
0x06
-
-
-
-
-
-
Number of entries.
Get gross weight GG command.
Get net weight GN command.
Get tare weight GT command.
Get dosed weight GD command
(FT1 & FT3 only)
.
Get dosed tare weight DT command
(FT1 & FT3 only)
.
Get average weight GA command.
0x2004
0x00
0x01
0x02
0x03
0x04
0x05
0x0060
0x0061
0x0062
0x0063
0x0064
0x0065
Number of entries.
Save Analog.
Save Calibration.
Save General set-up.
Save Dosed.
Save Set-points.
UI8
UI8
UI8
UI8
UI8
UI8
ro
wo
wo
wo
wo
wo
0x05
-
-
-
-
-
Number of parameters.
Save analog output settings AS command.
Save calibration settings (TAC protected) CS command.
Save general set-up parameters WP command.
Save dosing setup parameters command SD command
(FT1 & FT3 only)
.
Save set-point parameters SS command.
0x2005
0x00
0x01
0x02
0x03
0x0080
0x0081
0x0082
0x0083
Number of entries.
Start command.
Stop command.
Trigger command.
UI8
UI8
UI8
UI8
ro
wo
wo
wo
0x03
-
-
-
Number of entries.
Start Cycle SC command
(FT1 & FT3 only)
.
Abort Cycle AC command
(FT1 & FT3 only)
.
Trigger TR command.
0x2006
0x00
0x01
0x02
0x03
0x00A0
0x00A1
0x00A2
0x00A3
Number of entries.
Not used.
Factory Default.
Command byte.
UI8
UI8
UI8
UI8
ro
wo
wo
wo
0x03
-
-
-
Number of entries.
Not used.
Set factory default values (TAC protected) FD command.
See [Cyclic data] ShortCmd.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flintec DAD143 Series and is the answer not in the manual?

Flintec DAD143 Series Specifications

General IconGeneral
BrandFlintec
ModelDAD143 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals