5522A
Operators Manual
6-4
Table 6-2. Error Mode Commands
Command Description
EDIT
Sets the edit field. PRI is specified for the output value in single output functions and the
primary output value in dual output functions.
EDIT? Returns the edit field setting.
ERR_REF Selects the error reference source.
ERR_REF? Returns the presently selected error reference source.
ERR_UNIT Chooses how UUT error is shown.
ERR_UNIT? Returns presently selected value of ERR_UNIT.
INCR
Increments or decrements the output (as selected by the edit field) and enters error
mode, the same as using the output adjustment knob in local operation.
MULT Multiplies the reference magnitude (as selected by the edit field).
NEWREF
Sets the reference value to be the present Calibrator output value, the same as pressing
the NEW REF key in local operation.
OLDREF
Sets the Calibrator output to the previously programmed reference value, the same as
pressing the ENTER key in local operation.
OUT_ERR? Returns the UUT error computed after shifting the output with the INCR command.
REFOUT?
Returns the value of the reference, which is the output values of the Calibrator the last
time a new reference was established with an OUT, NEWREF, or MULT.
EDIT
Sets the edit field. PRI is specified for the output value in single output functions and the
primary output value in dual output functions.
Table 6-3. External Connection Commands
Command Description
CUR_POST
Selects the active binding posts for current output. This applies to current and power
outputs.
CUR_POST? Returns the active binding posts for current output.
EARTH Connects or disconnects the internal guard shield from earth (chassis) ground.
EARTH?
Returns whether the internal guard shield is connected or disconnected from earth
(chassis) ground.
EXTGUARD Connects or disconnects the internal guard shield from the LO binding post.
EXTGUARD?
Returns whether the internal guard shields are connected or disconnected from earth
(chassis) ground.
LOWS? Returns whether or not the low terminals are internally open or tied together.
LOWS
Selects whether or not the low terminals are internally open or tied together for dual
outputs.
PRES_UNIT Sets the pressure display units.
PRES_UNIT? Returns the pressure display units.
RTD_TYPE Sets the Resistance Temperature Detector (RTD) type.