If INT is returned, the reference temperature return is 0 unless you are in a thermocouple
mode of operation and the Calibrator is in Operate.
Responses: INT, <value of reference temperature>,CEL (or FAR)
EXT, <value of reference temperature>,CEL (or FAR)
Example: TC_REF? returns INT,2.988E+01,CEL
Returns Internal, 29.88, Celsius, when the thermocouple reference is internal and at
29.88 °C. (If the temperature return for the internal reference is 0 (0.00E+00), the
Calibrator is not in Operate, and/or the Calibrator is not in a thermocouple mode.)
TC_TYPE
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xxx
(Thermocouple Type command) Sets the Thermocouple (TC) temperature sensor type.
The TC type is used when the output is set to a temperature value with the OUT command
and the temperature sensor type is set to TC with the TSENS_TYPE command. When the
thermocouple type is changed while simulating a temperature output, the temperature is
changed to 0 °C. Once set, the Calibrator retains the TC type until power off or reset.
Parameters: B (B-type thermocouple)
C (C-type thermocouple)
E (E-type thermocouple)
J (J-type thermocouple)
K (K-type thermocouple) (default)
N (N-type thermocouple)
R (R-type thermocouple)
S (S-type thermocouple)
T (T-type thermocouple)
X (10 μV/°C linear output)
Y (% relative humidity)
Z (1 mV/°C linear output)
Example: TC_TYPE J
Set the thermocouple type for simulating a temperature output to a J-type thermocouple.
TC_TYPE?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Thermocouple Type query) Returns the Thermocouple (TC) temperature sensor type.
When the thermocouple type is changed while simulating a temperature output, the
temperature is changed to 0 °C.
Responses: B (B-type thermocouple)
C (C-type thermocouple)
E (E-type thermocouple)
J (J-type thermocouple)
K (K-type thermocouple) (default)
N (N-type thermocouple)
R (R-type thermocouple)
S (S-type thermocouple)
T (T-type thermocouple)
X (10 μV/°C linear output)
Y (% relative humidity)
Z (1 mV/°C linear output)
Example: TC_TYPE? returns K