EasyManua.ls Logo

Fluke 5522A - Page 231

Fluke 5522A
426 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
Commands
6
6-43
Returns K when the thermocouple type for simulating a temperature output is a K-type
thermocouple.
TC_TYPE_D
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Thermocouple Type Default command) Sets the default thermocouple (TC) sensor type,
which is saved in the Calibrator non-volatile memory. (While saving configuration data
in the non-volatile memory, a period of about 2 seconds, the Calibrator does not respond
to remote commands.) The TC type is set to the default at power on and reset.
Responses: B (B-type thermocouple)
C (C-type thermocouple)
E (E-type thermocouple)
J (J-type thermocouple)
K (K-type thermocouple) (default)
N (N-type thermocouple)
R (R-type thermocouple)
S (S-type thermocouple)
T (T-type thermocouple)
X (10 μV/°C linear output)
Y (% relative humidity)
Z (1 mV/°C linear output)
Example: TC_TYPE_D J
Set the thermocouple type default to a type-J thermocouple.
TC_TYPE_D?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Thermocouple Type Default query) Returns the default thermocouple (TC) sensor type.
Responses: B (B-type thermocouple)
C (C-type thermocouple)
E (E-type thermocouple)
J (J-type thermocouple)
K (K-type thermocouple) (default)
N (N-type thermocouple)
R (R-type thermocouple)
S (S-type thermocouple)
T (T-type thermocouple)
X (10 μV/°C linear output)
Y (% relative humidity)
Z (1 mV/°C linear output)
Example: TC_TYPE_D? returns K
Returns K when the thermocouple type default is a type-K thermocouple.
TEMP_STD
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Temperature Degree Standard command) Selects the temperature standard ipts-68 (1968
International Provisional Temperature Standard) or its-90 (1990 International
Temperature Standard), which is saved in the Calibrator non-volatile memory. (While
saving configuration data in the non-volatile memory, a period of about 2 seconds, the
Calibrator does not respond to remote commands.) The default is its-90.
Parameters: IPTS_68
ITS_90

Table of Contents

Related product manuals