Maintenance
Performance Tests
7
7-9
Table 7-5. Verification Tests for DC Current (AUX) (cont.)
Range Output Lower Limit Upper Limit
2.99999 A 1.09000 A 1.08979 A 1.09021 A
2.99999 A -1.09000 A -1.09021 A -1.08979 A
2.99999 A 2.99000 A 2.98906 A 2.99094 A
2.99999 A -2.99000 A -2.99094 A -2.98906 A
20.5000 A 0.0000 A -0.0005 A 0.0005 A
20.5000 A 10.9000 A 10.8954 A 10.9046 A
20.5000 A -10.9000 A -10.9046 A -10.8954 A
20.5000 A 20.0000 A 19.9833 A 20.0168 A
20.5000 A -20.0000 A -20.0168 A -19.9833 A
Table 7-6. Verification Tests for Resistance
Range Output Lower Limit Upper Limit
10.9999 Ω 0.0000 Ω -0.0010 Ω 0.0010 Ω
10.9999 Ω 2.0000 Ω 1.9989 Ω 2.0011 Ω
10.9999 Ω 10.9000 Ω 10.8986 Ω 10.9014 Ω
32.9999 Ω 11.9000 Ω 11.8982 Ω 11.9018 Ω
32.9999 Ω 19.0000 Ω 18.9980 Ω 19.0020 Ω
32.9999 Ω 30.0000 Ω 29.9978 Ω 30.0023 Ω
109.9999 Ω 33.0000 Ω 32.9979 Ω 33.0021 Ω
109.9999 Ω 109.0000 Ω 108.9962 Ω 109.0038 Ω
329.9999 Ω 119.0000 Ω 118.9954 Ω 119.0046 Ω
329.9999 Ω 190.0000 Ω 189.9938 Ω 190.0062 Ω
329.9999 Ω 300.0000 Ω 299.9914 Ω 300.0086 Ω
1.099999 kΩ 0.330000 kΩ 0.329991 kΩ 0.330009 kΩ
1.099999 kΩ 1.090000 kΩ 1.089974 kΩ 1.090026 kΩ
3.299999 kΩ 1.190000 kΩ 1.189954 kΩ 1.190046 kΩ
3.299999 kΩ 1.900000 kΩ 1.899938 kΩ 1.900062 kΩ
3.299999 kΩ 3.000000 kΩ 2.999914 kΩ 3.000086 kΩ
10.99999 kΩ 3.30000 kΩ 3.29991 kΩ 3.30009 kΩ
10.99999 kΩ 10.90000 kΩ 10.89974 kΩ 10.90026 kΩ
32.99999 kΩ 11.90000 kΩ 11.89954 kΩ 11.90046 kΩ
32.99999 kΩ 19.00000 kΩ 18.99938 kΩ 19.00062 kΩ
32.99999 kΩ 30.00000 kΩ 29.99914 kΩ 30.00086 kΩ