Table 7-6. Verification Tests for Resistance (cont.)
Range Output Lower Limit Upper Limit
109.9999 kΩ 33.0000 kΩ 32.9991 kΩ 33.0009 kΩ
109.9999 kΩ 109.0000 kΩ 108.9974 kΩ 109.0026 kΩ
329.9999 kΩ 119.0000 kΩ 118.9950 kΩ 119.0050 kΩ
329.9999 kΩ 190.0000 kΩ 189.9933 kΩ 190.0068 kΩ
329.9999 kΩ 300.0000 kΩ 299.9905 kΩ 300.0095 kΩ
1.099999 MΩ 0.330000 MΩ 0.329990 MΩ 0.330010 MΩ
1.099999 MΩ 1.090000 MΩ 1.089971 MΩ 1.090029 MΩ
3.299999 MΩ 1.190000 MΩ 1.189922 MΩ 1.190078 MΩ
3.299999 MΩ 1.900000 MΩ 1.899894 MΩ 1.900106 MΩ
3.299999 MΩ 3.000000 MΩ 2.999850 MΩ 3.000150 MΩ
10.99999 MΩ 3.30000 MΩ 3.29959 MΩ 3.30041 MΩ
10.99999 MΩ 10.90000 MΩ 10.89875 MΩ 10.90125 MΩ
32.99999 MΩ 11.90000 MΩ 11.89512 MΩ 11.90488 MΩ
32.99999 MΩ 19.00000 MΩ 18.99370 MΩ 19.00630 MΩ
32.99999 MΩ 30.00000 MΩ 29.99150 MΩ 30.00850 MΩ
109.9999 MΩ 33.0000 MΩ 32.9838 MΩ 33.0162 MΩ
109.9999 MΩ 109.0000 MΩ 108.9534 MΩ 109.0466 MΩ
329.9999 MΩ 119.0000 MΩ 118.6025 MΩ 119.3975 MΩ
329.9999 MΩ 290.0000 MΩ 289.1750 MΩ 290.8250 MΩ
1100.000 MΩ 400.000 MΩ 394.700 MΩ 405.300 MΩ
1100.000 MΩ 640.000 MΩ 631.820 MΩ 648.180 MΩ
1100.000 MΩ 1090.000 MΩ 1076.420 MΩ 1103.580 MΩ
Table 7-7. Verification Tests for AC Voltage (Normal)
Range Output Frequency Lower Limit Upper Limit
32.999 mV 3.000 mV 45 Hz 2.994 mV 3.006 mV
32.999 mV 3.000 mV 10 kHz 2.994 mV 3.006 mV
32.999 mV 30.000 mV 9.5 Hz 28.335 mV 31.665 mV
32.999 mV 30.000 mV 10 Hz 29.976 mV 30.024 mV
32.999 mV 30.000 mV 45 Hz 29.990 mV 30.010 mV
32.999 mV 30.000 mV 1 kHz 29.990 mV 30.010 mV
32.999 mV 30.000 mV 10 kHz 29.990 mV 30.010 mV
32.999 mV 30.000 mV 20 kHz 29.989 mV 30.011 mV