EasyManua.ls Logo

Fluke 5522A - Page 256

Fluke 5522A
426 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5522A
Operators Manual
7-18
Table 7-13. Verification Tests for Phase Accuracy, V and V
Range,
Normal
Output, V
Output,
Normal V
Frequency
Range,
AUX
Output
Output,
AUX
Phase °
Lower
Limit°
Upper
Limit °
65 Hz -0.10 0.10
400 Hz -0.25 0.25
1 kHz -0.50 0.50
5 kHz -2.50 2.50
10 kHz -5.00 5.00
30 kHz
0
-10.00 10.00
65 Hz 59.90 60.10
400 Hz 59.75 60.25
1 kHz 59.50 60.50
5 kHz 57.50 62.50
10 kHz 55.00 65.00
30 kHz
60
50.00 70.00
65 Hz 89.90 90.10
400 Hz 89.75 90.25
1 kHz 89.50 90.50
5 kHz 87.50 92.50
10 kHz 85.00 95.00
3.29999 3.00000
30 kHz 80.00 100.00
32.9999 30.0000 65 Hz 89.90 90.10
329.999 50.000 65 Hz
3.29999 V 3.00000 V
90
89.80 90.10

Table of Contents