Maintenance
Performance Tests
7
7-19
Table 7-14. Verification Tests for Phase Accuracy, V and I
Range,
Normal
Output, V
Output,
Normal V
Frequency
Range,
AUX
Output
Output,
AUX
Phase
°
Lower
Limit °
Upper
Limit °
65 Hz 329.99 mA 300.00 mA -0.10 0.10
1 kHz 329.99 mA 300.00 mA -0.50 0.50
30.000 mV
30 kHz 329.99 mA 300.00 mA -10.00 10.00
200.000 mV 65 Hz 2099999 A 2.00000 A -0.10 0.10
65 Hz 20.5000 A 5.0000 A -0.10 0.10
50.000 mV
400 Hz 20.5000 A 5.0000 A
0
-0.25 0.25
30.000 mV 65 Hz 329.99 mA 300.00 mA 59.90 60.10
65 Hz 2.99999 A 2.00000 A 59.90 60.10
65 Hz 20.5000 A 20.0000 A 59.90 60.10
329.999 mV
200.000 mV
400 Hz 20.5000 A 20.0000 A
60
59.75 60.25
65 Hz 329.99 mA 300.00 mA -0.10 0.10
65 Hz 2.99999 A 2.00000 A -0.10 0.10
65 Hz 20.5000 A 5.0000 A -0.10 0.10
400 Hz 20.5000 A 5.0000 A
0
-0.25 0.25
65 Hz 329.99 mA 300.00 mA 89.90 90.10
65 Hz 2.99999 A 2.00000 A 89.90 90.10
65 Hz 20.5000 A 20.0000 A 89.90 90.10
32.999 mV 3.3000 V
400 Hz 20.5000 A 20.0000 A
90
89.75 90.25
65 Hz 329.99 mA 300.00 mA -0.10 0.10
65 Hz 2.99999 A 2.00000 A -0.10 0.100
65 Hz 20.5000 A 5.0000 A -0.10 0.10
400 Hz 20.5000 A 5.0000 A
0
-0.25 0.25
65 Hz 329.99 mA 300.00 mA 89.90 90.10
65 Hz 2.99999 A 2.00000 A 89.90 90.10
65 Hz 20.5000 A 20.0000 A 89.90 90.10
329.999 V 33.000 V
400 Hz 20.5000 A 20.0000 A
90
89.75 90.25