EasyManua.ls Logo

Fluke 5522A - Page 257

Fluke 5522A
426 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
Performance Tests
7
7-19
Table 7-14. Verification Tests for Phase Accuracy, V and I
Range,
Normal
Output, V
Output,
Normal V
Frequency
Range,
AUX
Output
Output,
AUX
Phase
°
Lower
Limit °
Upper
Limit °
65 Hz 329.99 mA 300.00 mA -0.10 0.10
1 kHz 329.99 mA 300.00 mA -0.50 0.50
30.000 mV
30 kHz 329.99 mA 300.00 mA -10.00 10.00
200.000 mV 65 Hz 2099999 A 2.00000 A -0.10 0.10
65 Hz 20.5000 A 5.0000 A -0.10 0.10
50.000 mV
400 Hz 20.5000 A 5.0000 A
0
-0.25 0.25
30.000 mV 65 Hz 329.99 mA 300.00 mA 59.90 60.10
65 Hz 2.99999 A 2.00000 A 59.90 60.10
65 Hz 20.5000 A 20.0000 A 59.90 60.10
329.999 mV
200.000 mV
400 Hz 20.5000 A 20.0000 A
60
59.75 60.25
65 Hz 329.99 mA 300.00 mA -0.10 0.10
65 Hz 2.99999 A 2.00000 A -0.10 0.10
65 Hz 20.5000 A 5.0000 A -0.10 0.10
400 Hz 20.5000 A 5.0000 A
0
-0.25 0.25
65 Hz 329.99 mA 300.00 mA 89.90 90.10
65 Hz 2.99999 A 2.00000 A 89.90 90.10
65 Hz 20.5000 A 20.0000 A 89.90 90.10
32.999 mV 3.3000 V
400 Hz 20.5000 A 20.0000 A
90
89.75 90.25
65 Hz 329.99 mA 300.00 mA -0.10 0.10
65 Hz 2.99999 A 2.00000 A -0.10 0.100
65 Hz 20.5000 A 5.0000 A -0.10 0.10
400 Hz 20.5000 A 5.0000 A
0
-0.25 0.25
65 Hz 329.99 mA 300.00 mA 89.90 90.10
65 Hz 2.99999 A 2.00000 A 89.90 90.10
65 Hz 20.5000 A 20.0000 A 89.90 90.10
329.999 V 33.000 V
400 Hz 20.5000 A 20.0000 A
90
89.75 90.25

Table of Contents