Warning and Error Messages
B
B-3
Table B-1. 810 Vibration Tester Warning and Error Messages (cont.)
Display Message Description
Diagnosis failed The Tester cannot diagnose the measurement data. Make the measurement again. If the
problem continues, contact the Fluke Service Center for technical support.
External SD card not
present
This message shows during a self-test. External SD memory card is missing or the
installation is incorrect. See “Installing SD Memory Card” for the procedure.
External SD card failure This message shows during a self-test. SD card is faulty or corrupted.
Failed to perform self test This message shows during a self-test. The Tester is defective, contact the Fluke Service
Center for technical support.
Calibration past due. Last
calibration done on <Date>
Calibrate the Sensor. Contact the Fluke Service Center for technical support.
Set today’s date The current date in the Tester is before the calibration date. Set the Tester to the current
date. See “Tester Settings.”
Battery low, please re-
charge
Connect the Tester battery to the ac adapter, connect into a power source, and charge the
battery. See “How to Charge the Battery.”
Sensor failure This message shows during the self-test. The Sensor is defective. Contact the Fluke Service
Center for technical support.
Tachometer failure This message shows during the self-test. The Tachometer is defective. Contact the Fluke
Service Center for technical support.