LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–168
REV 9
6–2–62 CNTIF Assy and CNTIF2 Assy (FRU No. 502)
Time Required
30 Minutes
Tool Required
Screwdriver
Hexagonal box wrench
Procedure
Refer to ILLUSTRATION 6–86.
1. Turn OFF the system.
2. Remove the Rear CONN Panel Assy (FRU 501). Refer to 6–2–61 on page 6–166.
3. Unscrew six hexagonal studs (
1
6
) using box wrench.
4. Remove the CNTIF Assy or the CNTIF2 Assy.
CNTIF(2) ASSY DISASSEMBLY
ILLUSTRATION 6–86
Note
The CNTIF2 Assy includes the Eternet port for the optional DICOM network connection.