GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–188
REV 9
6–2–73 MODD (FRU No. 520), MODD Fan (FRU No. 521), MODD Holder Assy (FRU No. 522) (continued)
Emergency Ejection
When your disk is not mounted and the eject button does not function, eject the disk following the procedure below:
CAUTION
Use the emergency ejection procedure only in emergency situations. If this procedure is used
while a MO disk is inserted, data in the disk can be destroyed.
1. Turn the system power OFF.
2. Unfold and straighten one end of a large wire paper clip.
3. Insert the straightened end of the clip into the emergency eject hole and press firmly.
This will eject the disk.
EMERGENCY EJECTION
ILLUSTRATION 6–104