GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–189
REV 9
6–2–74 HDD Assy (FRU No. 523), HDD Holder Assy (FRU No. 524)
Time Required
15 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–105.
1. Make sure the customer data is saved on a MO disk. Refer to 4–3–3 (D), USER DATA BACK UP on page 4–22.
2. Turn OFF the system.
3. Remove the Keyboard Panel Assy (FRU 400). Refer to 6–2–49 on page 6–147.
4. Remove the HDDB Assy (FRU 522). Refer to 6–2–75 on page 6–195.
5. Unscrew two screws (
1
and
2
).
6. Remove the HDD Holder Assy.
7. Unscrew four screws (
3
–
6
).
8. Remove the HDD (FRU 523) out of the HDD Holder Assy (FRU 524).
Note
After replacing a HDD Assy with new one, the following works have to be done:
a. Set up the system preset parameters without regard to current settings. For set-up procedures, refer to
LOGIQ 400 Proprietary Manual, 2127662.
a–1. Set the Country/Zone properly. It takes approximately 40 minutes to complete this step.
a–2. Set other three parameters, Video Signal, Power Frequency and Modem Kind.
a–3. Reset the system using System Reset function of the Service Menu. Refer to LOGIQ 400 Proprietary
Manual, 2127662.
b. Install the customer data into new HDD using the back-up disk(s) after resetting the system. Refer to 4–3–3
(D), USER DATA BACK UP on page 4–22.
c. Set the User Option properly. To set the User Option properly, it is necessary to record the status of the options
using the User Option Display before replacing the HDD. For the User Options Set–Up, refer to LOGIQ 400
Proprietary Manual, 2127662.