GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–190
REV 9
6–2–74 HDD Assy (FRU No. 523), HDD Holder Assy (FRU No. 524) (continued)
Note
In the case of the B/W monitor console, it is necessary to change the settings of Option Set and Sys-
tem Parameters after replacing a HDD Assy and setting the Country/Zone. If the required changes
are not done properly, the system will not be functioning properly and the system error messages are
displayed on the CRT monitor screen frequently because the system is initialized as a CFM console.
The required changes are:
(1) The following items in the Option Set Menu shall be set to the “Not Installed”.
02 PW Doppler
03 Color Flow Mapping
04 B Color
05 Dop/CFM for Intravaginal Probe
06 Segmented Doppler
07 Real Time Auto Doppler Trace
17 Heading VCR Playback
(2) The following parameters in the System Parameter shall be set to the “White”.
Col Sys Inf
Col ImagCrs
Col Scale
Col Comment
Col Mes Crs
Col Applcat
Col Moving
Col Aux
Col DopTrace
For more details regarding the setting procedures, refer to LOGIQ 400 Proprietary Manual,
2127662 and User Manual.