GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–191
REV 9
6–2–74 HDD Assy (FRU No. 523), HDD Holder Assy (FRU No. 524) (continued)
FRU523
2.5–inch HDD (includ-
ing HDD Holder Assy
FRU524
HDD Holder
Assy
HDD ASSY AND HDD HOLDER ASSY DISASSEMBLY
ILLUSTRATION 6–105
3.5–Inch HDD Disassembly
Refer to ILLUSTRATION 6–106.
1. Make sure the customer data is saved on MO disk. Refer to 4–3–3 (D), USER DATA BACK UP on page 4–22.
2. Turn OFF the system.
3. Remove the Keyboard Panel Assy (FRU 400). Refer to 6–2–49 on page 6–147.
4. Unscrew two screws (
1
and
2
).
5. Pull out the HDD Holder2 Assy.
6. Disconnect three connectors, CB24N, CB5N, and CB45N. Refer to ILLUSTRATION 6–106.
7. Unscrew four screws (
3
6
).
8. Remove the HDD (FRU 523) out of the HDD Holder2 Assy (FRU 536).