GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–192
REV 9
6–2–74 HDD Assy (FRU No. 523), HDD Holder Assy (FRU No. 524) (continued)
3.5–Inch HDD Disassembly (continued)
Note
After replacing a HDD Assy with new one, the following works have to be done:
a. Set up the system preset parameters without regard to current settings. For set-up procedures, refer to
LOGIQ 400 Proprietary Manual, 2127662.
a–1. Set the Country/Zone properly. It takes approximately 40 minutes to complete this step.
a–2. Set other three parameters, Video Signal, Power Frequency and Modem Kind.
a–3. Reset the system using System Reset function of the Service Menu. Refer to LOGIQ 400 Proprietary
Manual, 2127662.
b. Install the customer data into new HDD using the back-up disk(s) after resetting the system. Refer to 4–3–3
(D), USER DATA BACK UP on page 4–22.
c. Set the User Option properly. To set the User Option properly, it is necessary to record the status of the options
using the User Option Display before replacing the HDD. For the User Options Set–Up, refer to LOGIQ 400
Proprietary Manual, 2127662.
HDD ASSY AND HDD HOLDER2 ASSY DISASSEMBLY
ILLUSTRATION 6–106