LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–195
REV 9
6–2–75 HDDB Assy (FRU No. 525)
Note
The HDDB Assy is not used for the console with the 3.5–inch HDD Assy.
Time Required
15 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–110.
1. Turn OFF the system.
2. Remove the Keyboard Panel Assy (FRU 400). Refer to 6–2–49 on page 6–147.
3. Disconnect three connectors (
1
3
).
4. Unscrew four screws (
4
7
).
5. Remove the HDDB Assy.
HDDB ASSY DISASSEMBLY
ILLUSTRATION 6–110