GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–196
REV 9
6–2–76 HDD LED Assy (FRU No. 526)
Note
The HDD LED Assy is not available on the console with the new type of 3.5–inch HDD Assy (Quan-
tum FireBall 1280S and Quantum FireBall EL).
Time Required
15 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–111.
1. Turn OFF the system.
2. Remove the Front Base Cover (FRU 305). Refer to 6–2–35 on page 6–131.
3. Remove the Front Cover (FRU 306). Refer to 6–2–36 on page 6–132.
4. Unscrew two screws (
1
and
2
).
5. Disconnect one connector (
3
).
6. Remove the HDD LED Assy.
HDD LED ASSY DISASSEMBLY
ILLUSTRATION 6–111