GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–197
REV 9
6–2–77 Power Switch Assy (FRU No. 527)
Time Required
15 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–112.
1. Turn OFF the system.
2. Remove the Front Base Cover (FRU 305). Refer to 6–2–35 on page 6–131.
3. Remove the Front Cover (FRU 306). Refer to 6–2–36 on page 6–132.
4. Unscrew two screws (
1
and
2
).
5. Disconnect one connector (
3
).
6. Remove the Power Switch Assy.
POWER SWITCH ASSY DISASSEMBLY
ILLUSTRATION 6–112