GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–245
REV 9
6–3–2 Replacement Procedures (continued)
(A) F1 on TRDR Assy
1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction
46–015902).
2. Remove the Side Cover Left. Refer to 6–2–31 on page 6–127.
3. Unscrew five screws and remove the Nest Cover. Refer to 6–2–78 on page 6–198.
4. Pull the TRDR Assy out of slot 1. Refer to 6–2–78 on page 6–198.
5. Replace the defective fuse with new one. Make sure that only fuse with the required rated current and of same
type is used for replacement. For the removal of the F1 fuse, refer to ILLUSTRATION 6–136.
6. After replacing Fuse, re-install the TRDR Assy, Nest Cover, and Side Cover Left.
REMOVING FUSE (F501) ON TRDR ASSY
ILLUSTRATION 6–136