EasyManua.ls Logo

GE LOGIQ 400 - Page 376

GE LOGIQ 400
526 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–246
REV 9
6–3–2 Replacement Procedures (continued)
(B) F1 on CNTIF Assy
1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction
46–015902).
2. Remove the Rear Cover. Refer to 6–2–33 on page 6–129.
3. Remove the Rear CONN Panel Assy. Refer to 6–2–61 on page 6–166.
Note
It is not necessary to disconnect cables in order to replace the fuse.
4. Replace the defective fuse with new one. Make sure that only fuse with the required rated current and of same
type are used for replacement.
5. After replacing Fuse, re-install the Rear CONN Panel Assy and Rear Cover.
REMOVING FUSE (F1) ON CNTIF ASSY
ILLUSTRATION 6–137

Table of Contents

Other manuals for GE LOGIQ 400

Related product manuals