GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–265
REV 9
6–6–4 Procedures (continued)
(A) Removing 2.5–inch HDD Assy, HDDB Assy and Cables (continued)
9. Disconnect the CB24M cable. Refer to ILLUSTRATION 6–146.
DISCONNECTING CB24 CABLE
ILLUSTRATION 6–146
10. Disconnect three cables, CB24M, CB5, and CB46, from the HDDB Assy. Refer to ILLUSTRATION 6–147.
REMOVING HDDB ASSY
ILLUSTRATION 6–147
11. Unscrew four screws and remove the HDDB Assy. Refer to ILLUSTRATION 6–147.