6–6–4 Procedures (continued)
(A) Removing 2.5–inch HDD Assy, HDDB Assy and Cables (continued)
12. Unscrew two screws and remove the 2.5–inch HDD Assy with HDD Holder Assy. Refer to
ILLUSTRATION 6–148.
REMOVING 2.5–INCH HDD ASSY
ILLUSTRATION 6–148
13. Disconnect the CB24M cable from the Mother Assy and Rear1 Assy and the CB5 cable from the Mother Assy.
Refer to ILLUSTRATION 6–149.
REMOVING CABLES
ILLUSTRATION 6–149
14. Remove the CB24M and CB5 cables.