6–7–4 Procedures (continued)
(B) Installing LV3 Unit (continued)
9. Connect one connector (
1
) as shown in ILLUSTRATION 6–168.
REAR VIEW OF LV3 UNIT (b)
ILLUSTRATION 6–168
10. Screw three (3) screws (
5
–
7
) to install the HV Unit. Refer to ILLUSTRATION 6–167 on page 6–282.
11. Connect one connector (
8
). Refer to ILLUSTRATION 6–167 on page 6–282.
12. Screw four (4) screws (
1
–
4
) to install the Right PS Cover. Refer to ILLUSTRATION 6–167 on page 6–282.