GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
TM
8–39 OPTIONS
REV 5
8–3–10 Operational Check–out (Continued)
g. Move the cursor to “NEXT” position by rotating the trackball and press the Set button four times to display the
fifth page (5/6).
SYSTEM PARAMETER SET-UP MENU (5/6)
ILLUSTRATION 8–44
Note
The “Quaf–Frm” format is available only for the following color printers:
UP–1850MD, UP–1850EPM, UP–2950MD, and UP–2850P
The “Quaf–Frmls” selection for the Color Printer Memory in the SYSTEM PARAMETER SETUP, is
not available for those printers. This function is valid only for the UP–3030MD printer.
h. Select the “Black and White Printer” from among Record1 B&W parameters. Refer to ILLUSTRATION 8–44.
i. Select the “Color Printer” from among Record1 Color parameters. Refer to ILLUSTRATION 8–44.
Note
Use Ellipse keys to select the Record1 and Record2 parameters. Other option is displayed by press-
ing the Ellipse keys when the current selection of Record1parameter or Record2 parameter is dis-
played in reverse video.
j. Move the cursor to “UP1850” for the Color Printer Type parameter and press Set to select it. Refer to
ILLUSTRATION 8–44.
k. Move the cursor to “Single” for the Color Printer Memory parameter and press Set to select it. Refer to
ILLUSTRATION 8–44.
l. Move the cursor to “RGB” for the Color Printer Signal parameter and press Set to select it. Refer to
ILLUSTRATION 8–44.