GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
FUNCTIONAL CHECKS
4–18
REV 5
4–3–3 Utility Menu (Continued)
(B) ERROR LOG DISPLAY
This loads and displays an error log file, or saves it into the MO (Magneto Optical) disk.
Type 2 and then press Return key when the Utility Menu is displayed. The Error Log Display is displayed as
shown in ILLUSTRATION 4–15.
ERROR LOG FILE DISPLAY
ILLUSTRATION 4–15
Procedure for Saving File
1. Type Y and then press Return key.
2. Insert an 128MB MO (Magneto Optical) disk (or 230MB disk for the conosle with software version 3.10 or later)
into the MOD drive of console and press any key.
Note
Use new initialized MO disk. It is impossible to save the error log file into the disk which another file is
already stored.
Note
The 230MB MO disk is available on the console with software version 3.10 or later.
During the time that the system is saving the error log file, “Copying now” is displayed on the CRT monitor.