4–3–3 Utility Menu (Continued)
(D) USER DATA BACK UP (Continued)
3. Input the file number you desire to save into the MO disk.
4. Press Return key. The system starts saving.
During the system is saving data, “Saving now” is displayed on the CRT monitor. After saving data, “Succeeded”
or “Failed” is displayed on the CRT monitor. Then, the message “Press Y to continue, Control + R to quit” is
displayed. Type Y and press Return key to continue to use this function. Press and hold Control, and press R to
quit this function.
Procedure for Loading User Data
1. Type 2 and then press the Return key. The message “Set a media to drive and then press any key” is dis-
played.
2. Insert an initialized 128MB MO (Magneto Optical) disk (or 230MB disk for the conosle with software version
3.10 or later) into the MOD drive of console and press the Return key. The User Data Menu is displayed on
the CRT monitor screen. Refer to ILLUSTRATION 4–18.
3. Input the file number you desire to load from the MO disk.
4. Press Return key. The system starts loading.
During the system is loading data, “Loading now” is displayed on the CRT monitor. After loading data, “Suc–
ceeded” or “Failed” is displayed on the CRT monitor. Then, the message “Press Y to continue, Control + R to
quit” is displayed. Type Y and press Return key to continue to use this function. Press and hold Control, and
press R to quit this function.
Note
After loading User Data, turn the system power OFF. Then, turn the system power ON again after