GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
FUNCTIONAL CHECKS
4–23
REV 5
4–3–3 Utility Menu (Continued)
(D) USER DATA BACK UP (Continued)
Type 4 and then press Return key when the Utility Menu is displayed. The User Data Back Up Menu is displayed
as shown in ILLUSTRATION 4–17.
USER DATA BACK UP MENU
ILLUSTRATION 4–17
Note
You cannot save two or more items into one MO disk. If you want to save two or more items of the User
Data Menu into one MO disk, Choose the item 09, All System & Application Data.
Procedure for Saving User Data
1. Type 1 and then press the Return key. The message “Set a media to drive and then press any key” is dis-
played.
2. Insert an initialized 128MB MO (Magneto Optical) disk (or 230MB disk for the conosle with software version
3.10 or later) into the MOD drive of console and press the Return key. The User Data Menu is displayed on the
CRT monitor screen as shown in ILLUSTRATION 4–18.
Note
Use new initialized MO disk. It is impossible to save any user data files into the disk which another file
is already stored.
USER DATA MENU
ILLUSTRATION 4–18