GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
FUNCTIONAL CHECKS
4–22
REV 5
4–3–3 Utility Menu (Continued)
(D) USER DATA BACK UP
This saves the user preset data into the MO disk. This also loads the user preset data stored in the MO disk.Sys-
tem Parameter, Probe Parameter, and the User Table into the external recording media.
Note
The user preset data is different depending on each category.
You can select the user preset data from among the following menu.
1. System Data
2. Application (Rad/Abdomen) Data
3. Application (Obstetrics) Data
4. Application (Gynecology) Data
5. Application (Cardiology) Data
6. Application (Vascular) Data
7. Application (Ulorogy) Data
8. Application (Small Parts) Data
9. All System & Application Data
System data includes:
System Preset Data File, User Define Data File*, VCR Registration Information
Data File*, VCR Patient Information Data File*, and Image Archive Search Data
File*
Each application data includes:
Application Preset Data File, Probe Preset Data File, Report Format Data File,
OB Table Data File, Fetal Trend Patient Data File*, and Cardiac Calc. Auto Se-
quence Data File*
The files with the symbol “*” listed above are saved only when the software options are installed and
any files are stored in the hard disk.
Note
The 230MB MO disk is available on the console with software version 3.10 or later.