FUNCTIONAL CHECKS
4–21
REV 5
4–3–3 Utility Menu (Continued)
(C) TROUBLE IMAGE RECORD & DISPLAY (Continued)
Procedure for Displaying Trouble Images
1. Type 3 and then press Return key. The Trouble Image File Menu will be displayed on the CRT monitor screen.
Refer to ILLUSTRATION 4–16.
2. Input the file number you desire to display on the CRT monitor screen.
3. Press Return key.
Press and hold the Control key and press Q to display the Trouble Image Menu on the CRT monitor screen.