4–3–3 Utility Menu (Continued)
(C) TROUBLE IMAGE RECORD & DISPLAY (Continued)
Procedure for Saving Trouble Images
1. Type 1 and then press Return key. The message “Set a media to drive and then press any key.” is displayed.
2. Set an 128MB MO (Magneto Optical) disk (or 230MB disk for the conosle with software version 3.10 or later)
into the MOD drive and press any key. The Trouble Image File Menu will be displayed on the CRT monitor
screen. Refer to ILLUSTRATION 4–16.
Note
Use new initialized MO disk. It is impossible to save any trouble image files into the disk which another
file is already stored.
Note
The 230MB MO disk is available on the console with software version 3.10 or later.
3. Input the file number you desire to save into the MO disk.
4. Press Return key.
During the system is saving the file, “Saving now” is displayed on the CRT monitor. After saving the file, “Suc–
ceeded” or “Failed” is displayed on the CRT monitor. Then, the message “Press Y to continue, Control + R to
quit” is displayed. Type Y and press Return to continue to use this function. Press and hold the Control key, and
press R to quit this function.
Procedure for Loading Trouble Images
1. Type 2 and then press Return key. The message “Set a media to drive and then press any key.” is displayed.
2. Set an 128MB MO (Magneto Optical) disk (or 230MB disk for the conosle with software version 3.10 or later)
into the MOD drive and press any key. The Trouble Image File Menu will be displayed on the CRT monitor
screen. Refer to ILLUSTRATION 4–16.
Note
The 230MB MO disk is available on the console with software version 3.10 or later.
3. Input the file number you desire to load from the MO disk.
4. Press Return key.
During the system is loading the file, “Loading now” is displayed on the CRT monitor. After saving the file, “Suc–
ceeded” or “Failed” is displayed on the CRT monitor. Then, the message “Press Y to continue, Control + R to
quit” is displayed. Type Y and press Return to continue to use this function. Press and hold the Control key, and
press R to quit this function.