DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
4 - 16 Section 4-2 - Functional checks
Recommended Test Phantoms
GE recommends the RMI 430GS phantom is preferred, but it is not required. It is the most current
phantom recommended to our field service personnel and provides the necessary targets and extended
life necessary for consistent Ultrasound system testing.
Figure 4-7 Performance tests
2
4321
4
6
8
10
12
14
16
RMI 403GS
Lateral Distance
Measurement
Functiona
l
Resolutio
n
Contrast
Resolution
&
Gray Scale
P
hotography
Lateral
Resolution
Axial
Resolution
Axial Distance
Measurement
Penetration