DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
3 - 24 Section 3-4 - Completing the setup
3-4-8 Password Policies
To create password policies:
1.) Navigate to Utility > Admin > Logon.
2.) Select Enable Password Policies. The system prompts the user for a password at logon.
3.) Under Password Policies, use the drop down selection beside each feature to set the rules for
passwords.
4.) Click Save.
Figure 3-10 Password Policies Window