DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 1 Introduction 1 - 21
1-4-3 Mechanical safety
WARNINGWARNING
While the software install procedure is designed to preserve data, you should save any
patient data, images, and system setups to a media or hardcopy before doing a
software upgrade.
WARNINGWARNING
PRIOR TO ELEVATING THE LOGIQ E10, VERIFY THE FOLLOWING:
- KEYBOARD IS LOCKED IN ITS LOWEST POSITION
- FRONT BRAKE IS LOCKED SO THE LOGIQ E10 IS UNABLE TO SWIVEL
- REAR BRAKES ARE IN THE LOCKED POSITION
WARNINGWARNING
WHEN THE LOGIQ E10 IS RAISED FOR A REPAIR OR MOVED ALONG ANY INCLINE,
USE EXTREME CAUTION SINCE IT MAY BECOME UNSTABLE AND TIP OVER.
WARNINGWARNING
ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN
EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND
PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR
DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN
SERIOUS INJURY AND EQUIPMENT DAMAGE.
WARNINGWARNING
NEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR. EVEN IF IT LOOKS OK, IT
MAY BE DAMAGED.
WARNINGWARNING
REMEMBER: If the front caster swivel lock is engaged for transportation, pressing the
release pedal once disengages the swivel lock. You must depress the release pedal a
second time to engage the brake.
WARNINGWARNING
The LOGIQ E10 should NOT be moved with the Operator I/O Panel extended. Move the
Operator I/O Panel to its centered and locked position. Lower the Operator I/O Panel as
much as possible before moving the LOGIQ E10.
CAUTION
BEFORE YOU MOVE OR TRANSPORT THE LOGIQ E10, MAKE SURE TO FIRMLY LOCK THE
MONITOR ARM AND FLIP DOWN THE MAIN MONITOR TO PREVENT DAMAGE TO THE
LOGIQ E10.
CAUTION
ALWAYS LOCK THE TOP/UPPER CONSOLE IN ITS PARKING (LOCKED) POSITION BEFORE
MOVING THE LOGIQ E10.
CAUTION
TO AVOID INJURY WHEN YOU MOVE THE MAIN MONITOR AND THE MONITOR ARM, DO NOT
PUT YOUR FINGER, HAND, OR OBJECT ON THE JOINT OF THE MAIN MONITOR OR THE
MONITOR ARM.