DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 106 Section 8-7 - Replacing Top Console Parts
8-7-5 Probe Holder Insert replacement
Table 8-115 Probe Holder Insert replacement Introduction
Introduction
The Probe Holder Inserts are soft rubber inserts, used to protect the Probes from scratches,
when stored on the LOGIQ E10. You can place the Probe Holder Inserts in any of the desired
places on the edge of the Operator Panel. The following types of inserts are available:
• Probe Holder Insert STD,
• Probe Holder Insert 3D,
• Probe Holder Softinsert Doppler.
Table 8-116 Manpower / Time and Tools
Manpower /
Time
Tools / PPE
One person /
15 minutes
No tools are required.
Table 8-117 Preparations
Preparations - you must perform the following steps
1. Disconnect and remove all Probes.
2. Store Probes in a safe place.