DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
4 - 18 Section 4-2 - Functional checks
B-Mode Checks
For information on the Ultrasound system’s different modes as well as expected results, refer to the
appropriate version of the LOGIQ E10 Basic User Manual, Chapter 5, or the appropriate LOGIQ E10
Release Notes. See: Section 1-5 "Label locations" on page 1-26.
Introduction
The B-mode is the Ultrasound system’s default mode.
Depending on probe availability, choose the preset application shown below. The LOGIQ E10
probes are ONLY approved for use in the applications listed in the LOGIQ E10 Basic User Manual,
Chapter 11, or the appropriate LOGIQ E10 Release Notes. See: Section 1-5 "Label locations" on
page 1-26.
IQC preset for Service is available.
Image Quality Check (IQC) is intended to facilitate Image Quality checks during Quality Assurance
Evaluations. Quality Assurance tests are used to determine whether a scanner is providing the same
level of performance year after year.
By using the same settings year after year, this ensures that the data collection is consistent,
independently of who performs the test.
This preset only includes fundamental settings for B-Mode.
Processing modes like SRI, Harmonics, etc., are turned off.
To do an Image Quality Check (IQC):
1.) Activate IQC via Utility -> Imaging Preset Manager -> Category (select the Category first).
2.) Click on the plus sign in front of IQC for Service.
3.) Assign IQC to a Touch Panel key by using the right arrow key.
4.) Map the IQC to the location you want it to appear on the Touch Panel.
5.) Select Model -> IQC.
Figure 4-8 B-mode Screen Example