5 - 22 Section 5-10 - Power distribution
5-10-2 Main Power Supply
General description
The Main Power Supply’s main task is to galvanically isolate the scanner from the on-site Mains Power
System and to supply the various internal subsystems with DC power.
Power from the wall outlet (100 to 240 VAC, 50/60 Hz) is connected to the Main Power Supply. The
Main Power Supply delivers the needed voltages to the rest of the LOGIQ E10:
• Printer (24 VDC)
• Front End Rack (DC power with several voltages)
• Front End Rack (TSV1 and TSV2 for the transmitters)
• Front End Rack (PMXVOUT for the probe channel multiplexers)
• Back End
An inrush current limiter reduces the peak input current when the LOGIQ E10 is switched on. An EMI
filter helps to reduce EMI to acceptable levels.
The mains cord has plugs in both ends. A female plug connects to the scanner and a male plug to the
wall outlet.
Fuses are located inside the power supply (only to be replaced by the manufacturer).