DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 6 Service Adjustments 6 - 17
Section 6-5
Adjust time-out for DICOM servers
If you are experiencing problems with slow responses from DICOM servers, increase the time-out in the
DICOM server properties dialog. (Utility -> Connectivity -> Service -> Properties -> Maximum
Retries and Timeout.)
Problems with slow responses may result in images being re-sent automatically and low transfer rates.
The retry settings can be used to make jobs retry on bad networks.
When portable (off-line), use minimum time-out and no retries or it will affect shutdown speed.
Section 6-6
Volume Navigation Calibration Procedure
There is no calibration available in the field. If the registration seems to be out of calibration, reloading
the software will restore the factory settings. If still out of calibration, replace hardware.