DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7 - 23
To change the password:
1.) Navigate to Utility (second page) > Service > Utilities > Change Password.
2.) Under User Type, select the user.
NOTE: Before changing the GEService password (the default is SvcForward123$), make sure the
LOGIQ E10 is connected to the network and the agent is configured. The GEService password
is used to perform portions of remote service. If the password is changed and the system
information is not updated, it may slow down remote service. Both file transfer and SSH depend
on the GEService password.
3.) In New Password and Confirm Password, enter the new password.
4.) Click Update Password.
5.) When a SVCService user password has been changed, reboot the LOGIQ E10 to reflect the
password change.
For more information, see:
• "Utilities" on page 7-19
CAUTION
RISK OF LOSING DATA.
GE WILL NOT BE ABLE TO RECOVER OR RESET CHANGED PASSWORDS.
SECURELY RECORD THE NEW PASSWORD.

Table of Contents

Other manuals for GE LOGIQ E10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E10 and is the answer not in the manual?

GE LOGIQ E10 Specifications

General IconGeneral
TouchscreenYes
3D/4D ImagingYes
ApplicationsVascular, Cardiac, Musculoskeletal, OB/GYN
Image StorageUSB, Network
ConnectivityDICOM, Ethernet, Wireless

Related product manuals