DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 110 Section 8-7 - Replacing Top Console Parts
Calibration and adjustments
No calibrations or adjustments are needed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Power up the system to verify that it operates as intended.
2.) Operate all knobs and verify that they don’t slip when operated.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
Table 8-125 OP Knobs replacement Functional Checks
See:
Section
Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 System CFM and PWD Checks
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN