DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 8 Replacement Procedures 8 - 161
Palm Rest Cover replacement
Table 8-187 Palm Rest Cover replacement
Steps Corresponding Graphic
1. Remove the six OP Knobs.
NOTICE
Failure to remove the six OP Knobs first, could cause damage to the knob shafts.
2. Remove the Upper OP.
3. Remove the two screws from under the
Palm Rest, using a #2 Phillips screwdriver.
Remove the Palm Rest Cover.
4. Make sure the XYZ Buttons cable runs along the Lower OP cable channel, as shown so the cable
is not pinched when the Palm Rest is replaced.
Install the replacement cover and screws.
These screws are screwed into plastic. Use care during installation, DO NOT overtighten.
5. Re-install Lower OP, Upper OP and all Covers removed.