DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 414 Section 8-16 - Options replacement
CW Option removal
For location of the Option in the LOGIQ E10, see: Table 8-401 "LOGIQ E10 Option Location" on page
8-408.
Table 8-409 CW Option removal
Steps Corresponding Graphic
1.
Pull the CW Option cable straight out to
disconnect from the Backplane.
2.
Loosen the two Phillips screws, using a #2
Phillips screwdriver, but do not remove.
Lift the CW Connector, Card Rack Bracket
straight up to remove.