DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
10 - 6 Section 10-3 - Maintenance Task Schedule
Surface Probe Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Endocavity Probe
Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Measurement Accuracy
Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Probe/Phantom Checks
Check axial and lateral resolution (see Basic
User Manual Technical specifications).
Check Gain and TGC changes, vary the
focus and check reaction on screen. Check
deviation of brightness in the US-Image
(missing elements / probe cable defect).
Probe must be coupled as this test.
•
Also after corrective
maintenance or as required by
your facilities QA program.
Functional Checks of all
probes section 10-5-3
on page 10-9
Check general functions and image
appearance at human body with all
available Modes.
•
Also after corrective
maintenance or as required by
your facilities QA program.
Table 10-2 Customer Care Schedule
Item Service at Indicated Time D W M A Notes