System Configuration
2028451-182B MAC™ 1600 9-17
Most treadmill and ergometer protocols consist of three pre-defined
phases: Pretest, Exercise, and Recovery. Each phase can include multiple
stages, which define the parameters of the test. The parameters differ
slightly depending on the device, as seen in the following table.
Protocols for Master’s Step devices consist of only Stage and Stage Time.
You can modify the pre-defined protocols to create custom protocols. Use
the following instructions to create a custom protocol.
1. From the MAC 1600 Main Menu, press F5 (System Configuration) >
F4 (Stress ECG Setup).
The Stress ECG Setup window opens.
2. Press F4 (Page Down).
The second page opens.
Ergometers WHO WHO50 WHO75
HOLLMANN BAL STD.FRANCE
MODWHO CONCONI
Masters Step SINGLE DOUBLE TRIPLE
Parameter Treadmill Ergometer Comment
Stage Y Y The stage name.
Stage Time Y Y The stage duration, in minutes.
Speed Y N The treadmill speed in kilometers or
miles per hour, depending on the
Country Setup.
Grade [%] Y N The percentage of increase in the
treadmill’s elevation.
Basic Load [W] N Y The load at which the ergometer
operates, in watts.
Store Median First Y Y The interval at which the first median
reading is stored.
Store Median
Repeat
Y Y The interval at which a subsequent
median reading is stored.
BP First Y Y The interval at which the first blood
pressure reading is stored.
BP Repeat Y Y The interval at which subsequent
blood pressure readings are stored.