9-18 MAC™ 1600 2028451-182B
System Configuration
3. Select the Edit Protocols button and press either Enter or the
trimpad button.
For treadmills and ergometers, the Select Protocol window opens to
display applicable protocols. Perform step 4 through step 16.
For Master’s Step devices, the Edit Master Step Post-Exercise
window opens to display the display the post-exercise stages.
Perform step 8 through step 12.
4. Press F2 (Add).
A list of templates opens.
5. Select the template to base the new protocol on.
The templates are based on the existing protocols. An additional
Empty Protocol is also available.
6. Press F6 (OK).
The Add Protocol window opens.
7. Type a name for the new protocol and press F6 (OK).
The Protocol window opens with all the stages from the template.
You can now add, edit, or delete stages.
8. To add a stage, do the following:
a. Select the stage that will precede the new stage.
b. Press F2 (Add Stage).
The selected stage is duplicated. Edit the duplicate stage as
appropriate. Refer to step 9.
9. To edit a stage, do the following:
a. Select the stage to edit.
b. Press F1 (Edit).
The Edit Stage window opens.
c. Modify the stage parameters as appropriate.
Refer to the table preceding these instructions for a description of
each parameter.
d. When you are done, press F6 (OK).
The Edit Stage window closes.
10. To delete a stage, do the following:
a. Select the stage to be deleted.
b. Press F3 (Delete Stage).
The selected stage is deleted.
11. To remove custom Master’s Step stages, press F4 (Factory Defaults).