Loading programs to flash memory
Theashmemorycanbeloadedwithnewwashprogramsornewsystemprograms.
LoadingnewwashprogramsrequirestheCS-1000program,whichcanbepurchased
from Getinge. Instructions are supplied with CS-1000. System programs are loaded with
Flashloader.
Note:
Always make a backup copy before starting work on updating system programs.
1. Connect a PC to the machine; see under Connecting a PC.
2. CheckthatthemachineisinSTANDBYmode.
3. Make a backup copy by starting CS 1000 and choosing:
Tools/PACS RAM/Upload To File...









V1544
V1577
4. Savethe*.prmleinyourchosenlocation.
The program will report an error during conversion. Disregard this. To check that
conversionwassuccessful,checkthesizeofthe*.prmle.Thesizeoftheleshould
be 84416 bytes.
5. Switch off the power to the machine with the main switch.
6. Change the jumper (X29) on the board for the PACS 300 control system from
Normal to Test.
X29

Table of Contents

Other manuals for Getinge 86 SERIES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 86 SERIES and is the answer not in the manual?

Getinge 86 SERIES Specifications

General IconGeneral
BrandGetinge
Model86 SERIES
CategoryCleaning Equipment
LanguageEnglish

Related product manuals