FCX4000-UM-251-9370 8-21
8 ELECTRICAL ADJUSTMENTS
(11) Similarly, adjust the 40 g-pen force as like as step 10.
PresstheENTERkeyifthemeasuredvalueiswithinthespecicationrange(48±4g).
Press the ENTER key if you have input the recorded value.
Thenextspeciedpenforceappears.
(12) Similarly, adjust the 130 g-pen force as like as step 10.
PresstheENTERkeyifthemeasuredvalueiswithinthespecicationrange(130±10g).
Press the ENTER key if you have input the recorded value.
Thenextspeciedpenforceappears.
(13) Similarly, adjust the 210 g-pen force as like as step 10.
PresstheENTERkeyifthemeasuredvalueiswithinthespecicationrange(210±20g).
Press the ENTER key if you have input the recorded value.
Thenextspeciedpenforceappears.
(14) Similarly, adjust the 515 g-pen force as like as step 10.
PresstheENTERkeyifthemeasuredvalueiswithinthespecicationrange(515±20g).
Press the ENTER key if you have input the recorded value.
Thenextspeciedpenforceappears.
(15) Similarly, adjust the 618 g-pen force as like as step 10.
PresstheENTERkeyifthemeasuredvalueiswithinthespecicationrange(618±20g).
Press the ENTER key if you have input the recorded value
(16) TheadjustmentmenusfortheNo.2penforcewillbedisplayedafteryouhavenishedtheadjustments
for the No.1 pen force.
Adjust the pen pressure for the No.2 pen force in the same way as you did for the No.1 pen force.
(17) Proceed to the adjustments for the 10 g to the 618 g.
(18) When the adjustments have been completed for the No.2 pen force, the following menu is displayed.
Whenthemeasuredpenforceisthespeciedvalue,presstheENTERkey.Thenextspeciedpenforce
appears on the LCD panel.
Repeat the above procedure in the following sequence.
Displayed
target force
No.1 pen pressure
Displayed
target force
No.2 pen pressure
10 10g±2g 10 10g±2g
48 48g±4g 48 48g±4g
130 130g±10g 130 130g±10g
210 210g±20g 210 210g±20g
515 515g±20g 515 515g±20g
618 618g±20g 618 618g±20g
(19) Press the ESC key to exit from the pen force adjustment menu.
(20) Record the adjusted values to the factory adjustment values label when the adjustment values are
changed.