FUNCTION MENU
159
IHVL(I range high, V range low)
ILVH(I range low, V range high)
IMVH(I range middle, V range high)
IHVH(I range high, V range high)
Setting
Sets the values corresponding to the
defined discharging mode (CC
mode in A, CR mode in mS and CP
mode in W).
SlewRate=
Sets the test rising slew rate in
mA/us (not adjustable for CP
mode).
SlewRate?
Sets the test falling slew rate in
mA/us (not adjustable for CP
mode).
Stop Volt
Sets the voltage at which the test
should be interrupted. The value
must be lower than the battery start
voltage.
Stop Time
Sets the time after which the test
should be interrupted (max value is
999h:59m:59s).
Stop AH
Sets the discharged energy rate at
which the test should be interrupted
(Max value is 9999.99Ah).
Datalog timer
Sets the time interval for data
capture. Up to 65,535 data can be
saved when running data logging
function. When logging data reaches
to the maximum amount, it wont be
saved and be ignored.
Panel operation
1. Press
FUNC
> Next Manu[F5]. > BATT[F2].