PEL-3000(H) User Manual
60
+CV Mode
+CV mode can be added to CC, CR and CP mode.
The +CV settings apply to all applicable modes.
1. Make sure the load is off.
2. Press
and select to Mode, I Range and
V Range.
3. Set the +CV voltage level. (You may need to
scroll down to the +CV setting)
Mode
CP + CV
I Range
H 8.75A
V Range
L 80V
Function
Dynamic
Configure
0.000
V
0.0000
A
0.00
W
+CV 5.500
Timer2 0.025
Timer1 0.025
LOADRS232
01/Oct/2018
V
ms
ms
+CV setting
The +CV settings apply to all the applicable operating
modes.
For example: The +CV settings made in CR mode will
be carried over to the +CV settings in CC and CP
mode.
+CV settings cannot be controlled with external
control. (The external control is not possible only in
+CV settings)
See page 168 for +CV settings with external control