PEL-3000(H) User Manual
186
Current Monitor Output
Description
The voltage output from the IMON OUT terminal
and from the IMON pin on the J3 connector is used
to represent the current input level.
The voltage range used to represent the full scale
current range from the IMON OUT terminal and
from the IMON pin on the J3 connector depends
on the current range settings.
Current
IMON V
Rated
Current
Monitor Output Range
Monitor Connector
Current Range
Monitor Output
Range
I MON OUT (BNC)
H, L
0 - 1V
M
0 - 0.1V
I MON (J3)
H, L
0 - 10V
M
0 - 1V
I MON OUT BNC
Connector
The IMON OUT BNC connector outputs a voltage
of 0 - 1V for the High and Low current ranges and
0 - 0.1V for the Middle current range. The
common potential is connected to the chassis
ground potential.
J1 Connector
The voltage across pins 2 and 3 outputs a voltage
of 0 -10V for the High and Low current ranges and
0 - 1V for the Middle current range. The common
potential is connected to A COM (negative load
terminal).