OPERATION
83
Display
8.653
V
50.006
mA
0.4327
W
CV B Value 80.000
CV A Value 80.000
LOADRS232
A Value
Fine
01/Oct/2017
V
V
OCP Alarm
Please Press Enter
To Clear Alarm
OCP
OCP indicator
Alarm message
when OCP is
set to Load Off
Mode
CV
I Range
L87.5mA
V Range
L 80V
Response
Fast
Configure
OPP
Description
For OPP, the PEL-3000(H) can be configured to
either limit the power or turn off the load.
The OPP levels can be set to 10% higher than the
rating power.
Operation
1. Press
Main
> Configure[F5] > Protection[F1]
and set the OPP Level and OPP Setting.
Range:
OPP Level: Rating power + 10%
OPP Setting: LIMIT, Load Off
Alarm
When OPP Setting is configured to Load Off, a
message will be displayed on the screen when
OPP is tripped. The Enter key must be pressed
to clear the alarm message.
When configured to LIMIT, OPP will be
displayed on the screen when the OPP is
tripped and the power will be limited to the
OPP Level setting.