For OPP, the PEL-3000(H) can be configured to
either limit the power or turn off the load.
The OPP levels can be set to 10% higher than the
rating power.
OPP Level: Rating power + 10%
OPP Setting: LIMIT, Load Off
When OPP Setting is configured to Load Off, a
message will be displayed on the screen when
OPP is tripped. The Enter key must be pressed
to clear the alarm message.
When configured to LIMIT, OPP will be
displayed on the screen when the OPP is
tripped and the power will be limited to the
OPP Level setting.